ayanjoshi2003's picture
Update README.md
518f748 verified
|
raw
history blame
45.1 kB
metadata
license: apache-2.0
language:
  - en
pipeline_tag: summarization
widget:
  - text: >-
      the need for magnetic resonance imaging ( mri ) in patients with an
      implanted pacemaker or implantable cardioverter - defibrillator ( icd ) is
      a growing clinical issue . it is estimated that as many as 75% of active
      cardiac device recipients will become indicated for mri . currently , the
      vast majority of such devices are contraindicated for use with an mri . in
      european heart rhythm association survey , published recently for non -
      mri - certified icds ( 0.51.5 t field strength ) , the totally
      subcutaneous icd ( s - icd ) system , an implantable defibrillator with no
      leads that touch the heart , has recently been demonstrated to be a safe
      and effective defibrillator option for patients at risk for sudden cardiac
      death . it provides shock therapy and post - shock pacing therapy , but no
      long - term bradycardia pacing . although it has been shown as an
      alternative to the standard transvenous icd , its compatibility with mri
      remains unclear . various types of clinical mri systems currently use a
      superconductive magnet that creates a static magnetic field strength ,
      typically 1.5 or 3 t. the use of mri with most pacemakers and icds is
      considered a contraindication due to potential hazards , including heating
      of the electrode that resides in or on the heart , damage to myocardium ,
      elevation of pacing thresholds , unintended induction of ventricular
      tachycardia ( vt ) or ventricular fibrillation ( vf ) , pacing inhibition
      , permanent device malfunction , and distortion of the mri scan . recently
      , mr - conditional. mr - conditional indicates a lack of known hazards in
      a specified mri environment with specified conditions of use . due to the
      variety of mri scanners and scanning protocols , it is not practical to
      test even a single device under all conditions . hence , mr - conditional
      labelling dictates that the device is safe for use under certain scanning
      conditions , as well as how the cardiac device should be programmed before
      an exposure to the magnetic field in a mri scanner . the literature ,
      although limited , provides some guidance for imaging patients with
      implanted pacemakers or icds that do not have mr - conditional labelling .
      this single - centre prospective non - controlled study describes the
      first use of mri in patients with an implanted s - icd . patients with
      implanted s - icd systems ( boston scientific sqrx model 1010 and q - trak
      model 3010 ) were enrolled for mri testing over a period of 18 months .
      the s - icd system implanted in this patient cohort was composed of a can
      implanted in a left mid - lateral pocket and a para - sternal subcutaneous
      electrode . the s - icd is currently not certified for use with an mri ;
      therefore , the ethics committee of homolka hospital , prague , czech
      republic approved our clinical study . patients with newly implanted s -
      icd systems ( < 6 weeks ) were excluded , and none of the patients had any
      intravascular leads . the patients were randomized for either a cardiac ,
      brain , cervical , or lumbar spinal scan . one of the subjects underwent
      an additional knee examination , due to reported chronic pain . a total of
      15 patients were enrolled into this study ( 12 males and three females ,
      aged 2283 years , mean 53 years . subjects in our cohort ( table 1 )
      underwent a total of 22 mri scans between 6 june 2012 and 24 december 2013
      . in total , five brain scans , three cardiac scans , 12 lumbar scans ,
      one knee , and one cervical spine scan were conducted ( table 2 ) .
      however , in one patient a minor disc protrusion was found , in other mri
      revealed stenosis of intervertebral foramen which was causing radicular
      pain of the nerve root l4 and based on this examination the patient was
      referred to ct - navigated periradicular therapy . table 1summary of
      patient anatomical data and scan locations , along with noted clinical
      eventsidagesexbmidgef , % indication for s - icdheating0164f20.5hcmp /
      vfs85secondary preventionnone0283m30.0post - mi / smvts post - catheter
      ablation/35secondary prevention ( post - transvenous icd
      extraction)none0331m25.3arvc / d / smvts68secondary preventionin -
      tolerable re - scanned0458m23.6post - mi / post - cabg30primary
      preventionnone0577m25.5post - mi30primary preventionnone0663m27.0post -
      mi30primary preventionnone0768m23.7post - mi / vfs / vts60secondary
      prevention post - transvenous icd extraction / svc
      occlusiontolerable0822m29.4brugada sy / vfs68secondary preventionin -
      tolerable re - scanned0959m27.1dcmp / vfs / post - mitral valve
      surgery/60secondary prev./post - transvenous icd
      extractionnone1041f24.6arvc / d70primary preventionnone1123f21.5lqts /
      vf60secondary preventionnone1266m36.9post - mi / vf / post -
      cabg50secondary prevention / post - repeat transvenous icd
      extractiontolerable1348m22.9dcmp(non - compaction)/vfs35secondary
      preventionnone1470m29systolic dysfunction of lv35primary
      preventionnone1526m33brugada sy65primary preventionnonehcmp , hypertrophic
      cardiomyopathy ; smvt , sustained monomorphic ventricular tachycardia ; mi
      , myocardial infarction ; arvc , arrhythmogenic right ventricular
      cardiomyopathy ; cabg , coronary artery by - pass graft ; lqts , long qt
      syndrom . table 2parmeters of s - icd and patient sensation during
      individual mri scansscan # idbody partheating sensationsshock zone (
      b.p.m.)condit . shock zone ( b.p.m.)bat % episode
      num.101brainnone2302101001202brainnone240220861303l spinein -
      tolerable240220831403brainnone240220831504brainnone220190691605l
      spinenone220210541706l spinenone240220681807l spinetolerable240220582908l
      spinein - tolerablenananana1008brainnonenananana1108l
      spinenone2302108411209heartnone2402208911310l
      spinenone2301807911410heartnonenananana1511heartnone2301909711612l
      spinetolerable2001709721712l spinenone2001709421813c
      spinenone23019010041913l spinenone23019010042014l
      spinenone2301908612115kneenone25021010012215l spinenone2502101001s - icd
      parameters acquired prior- and post - mri were without any change ,
      therefore only one value is presented.indices : na , not available ; l
      spine , lumbar spine ; c spine , cervical spine . summary of patient
      anatomical data and scan locations , along with noted clinical events hcmp
      , hypertrophic cardiomyopathy ; smvt , sustained monomorphic ventricular
      tachycardia ; mi , myocardial infarction ; arvc , arrhythmogenic right
      ventricular cardiomyopathy ; cabg , coronary artery by - pass graft ; lqts
      , long qt syndrom . parmeters of s - icd and patient sensation during
      individual mri scans s - icd parameters acquired prior- and post - mri
      were without any change , therefore only one value is presented . indices
      : na , not available ; l spine , lumbar spine ; c spine , cervical spine .
      studies were performed using a siemens avanto 1.5 t mri scanner ( vb17
      software , quantum gradient coils ) . all scans were run in normal
      operating mode , which is limited to 2 w / kg whole body averaged specific
      absorption rate ( sar ) . clinically relevant mri sequences were used for
      evaluation ( see table 3 ) . table 3types of pulse sequences typically
      used for imaging of respective anatomical areasscan locationscan
      sequencesflairdwiflashfsehastesestirtruefispbrainxxxxheartxxxxcervical
      spinexxkneexxxxlumbar spinexxflair , fluid attenuated inversion recovery ;
      dwi , diffusion weighted imaging ; flash , fast low angle shot ; fse ,
      fast spin echo ; haste , half acquisition single - shot turbo spin echo ;
      se , spin echo ; stir , short tau inversion recovery ; truefisp , true
      fast imaging with steady - state precession.fse sequence caused heating in
      subjects with a thermistor probe during lumbar spine examination ( see the
      text for details ) . types of pulse sequences typically used for imaging
      of respective anatomical areas flair , fluid attenuated inversion recovery
      ; dwi , diffusion weighted imaging ; flash , fast low angle shot ; fse ,
      fast spin echo ; haste , half acquisition single - shot turbo spin echo ;
      se , spin echo ; stir , short tau inversion recovery ; truefisp , true
      fast imaging with steady - state precession . fse sequence caused heating
      in subjects with a thermistor probe during lumbar spine examination ( see
      the text for details ) . patients were asked to report immediately any
      pain , torqueing movement , or heating sensation in the area of the pocket
      or the electrode by pressing an emergency bulb . furthermore , all
      patients were questioned immediately following the mri procedure to
      ascertain any discomfort in the vicinity of the can or electrode . pulse
      oximetry and standard lead electrocardiogram ( ecg ) if discomfort
      occurred , the patient was asked if the scan could be repeated at a later
      time using a revised scan sequence or the subject was again randomized for
      another anatomical area . since none of the components of the s - icd
      system are on or in the heart , heating near or around however , heating
      near the electrode or can with the s - icd system may still cause serious
      patient discomfort . therefore , along with education of subjects , each
      patient was instrumented by taping an oesophageal temperature probe ( beta
      - therm model g22k7mcd8 ) on the skin over the mid - lateral implant site
      to record any temperature excursions that might be correlated to patient
      symptoms of heating / discomfort near the pocket . to minimize the risk of
      inappropriate therapy , the s - icd system was programmed to therapy each
      s - icd system was evaluated prior to and immediately after the scan to
      verify proper functioning , including interrogation , sensing , and
      battery voltage . after the completion of the mri , long - term regular
      clinical follow - up and checking of the device were performed . patients
      with implanted s - icd systems ( boston scientific sqrx model 1010 and q -
      trak model 3010 ) were enrolled for mri testing over a period of 18 months
      . the s - icd system implanted in this patient cohort was composed of a
      can implanted in a left mid - lateral pocket and a para - sternal
      subcutaneous electrode . the s - icd is currently not certified for use
      with an mri ; therefore , the ethics committee of homolka hospital ,
      prague , czech republic approved our clinical study . patients with newly
      implanted s - icd systems ( < 6 weeks ) were excluded , and none of the
      patients had any intravascular leads . the patients were randomized for
      either a cardiac , brain , cervical , or lumbar spinal scan . one of the
      subjects underwent an additional knee examination , due to reported
      chronic pain . a total of 15 patients were enrolled into this study ( 12
      males and three females , aged 2283 years , mean 53 years . subjects in
      our cohort ( table 1 ) underwent a total of 22 mri scans between 6 june
      2012 and 24 december 2013 . in total , five brain scans , three cardiac
      scans , 12 lumbar scans , one knee , and one cervical spine scan were
      conducted ( table 2 ) . however , in one patient a minor disc protrusion
      was found , in other mri revealed stenosis of intervertebral foramen which
      was causing radicular pain of the nerve root l4 and based on this
      examination the patient was referred to ct - navigated periradicular
      therapy . table 1summary of patient anatomical data and scan locations ,
      along with noted clinical eventsidagesexbmidgef , % indication for s -
      icdheating0164f20.5hcmp / vfs85secondary preventionnone0283m30.0post - mi
      / smvts post - catheter ablation/35secondary prevention ( post -
      transvenous icd extraction)none0331m25.3arvc / d / smvts68secondary
      preventionin - tolerable re - scanned0458m23.6post - mi / post -
      cabg30primary preventionnone0577m25.5post - mi30primary
      preventionnone0663m27.0post - mi30primary preventionnone0768m23.7post - mi
      / vfs / vts60secondary prevention post - transvenous icd extraction / svc
      occlusiontolerable0822m29.4brugada sy / vfs68secondary preventionin -
      tolerable re - scanned0959m27.1dcmp / vfs / post - mitral valve
      surgery/60secondary prev./post - transvenous icd
      extractionnone1041f24.6arvc / d70primary preventionnone1123f21.5lqts /
      vf60secondary preventionnone1266m36.9post - mi / vf / post -
      cabg50secondary prevention / post - repeat transvenous icd
      extractiontolerable1348m22.9dcmp(non - compaction)/vfs35secondary
      preventionnone1470m29systolic dysfunction of lv35primary
      preventionnone1526m33brugada sy65primary preventionnonehcmp , hypertrophic
      cardiomyopathy ; smvt , sustained monomorphic ventricular tachycardia ; mi
      , myocardial infarction ; arvc , arrhythmogenic right ventricular
      cardiomyopathy ; cabg , coronary artery by - pass graft ; lqts , long qt
      syndrom . table 2parmeters of s - icd and patient sensation during
      individual mri scansscan # idbody partheating sensationsshock zone (
      b.p.m.)condit . shock zone ( b.p.m.)bat % episode
      num.101brainnone2302101001202brainnone240220861303l spinein -
      tolerable240220831403brainnone240220831504brainnone220190691605l
      spinenone220210541706l spinenone240220681807l spinetolerable240220582908l
      spinein - tolerablenananana1008brainnonenananana1108l
      spinenone2302108411209heartnone2402208911310l
      spinenone2301807911410heartnonenananana1511heartnone2301909711612l
      spinetolerable2001709721712l spinenone2001709421813c
      spinenone23019010041913l spinenone23019010042014l
      spinenone2301908612115kneenone25021010012215l spinenone2502101001s - icd
      parameters acquired prior- and post - mri were without any change ,
      therefore only one value is presented.indices : na , not available ; l
      spine , lumbar spine ; c spine , cervical spine . summary of patient
      anatomical data and scan locations , along with noted clinical events hcmp
      , hypertrophic cardiomyopathy ; smvt , sustained monomorphic ventricular
      tachycardia ; mi , myocardial infarction ; arvc , arrhythmogenic right
      ventricular cardiomyopathy ; cabg , coronary artery by - pass graft ; lqts
      , long qt syndrom . parmeters of s - icd and patient sensation during
      individual mri scans s - icd parameters acquired prior- and post - mri
      were without any change , therefore only one value is presented . indices
      : na , not available ; l spine , lumbar spine ; c spine , cervical spine .
      studies were performed using a siemens avanto 1.5 t mri scanner ( vb17
      software , quantum gradient coils ) . all scans were run in normal
      operating mode , which is limited to 2 w / kg whole body averaged specific
      absorption rate ( sar ) . clinically relevant mri sequences were used for
      evaluation ( see table 3 ) . table 3types of pulse sequences typically
      used for imaging of respective anatomical areasscan locationscan
      sequencesflairdwiflashfsehastesestirtruefispbrainxxxxheartxxxxcervical
      spinexxkneexxxxlumbar spinexxflair , fluid attenuated inversion recovery ;
      dwi , diffusion weighted imaging ; flash , fast low angle shot ; fse ,
      fast spin echo ; haste , half acquisition single - shot turbo spin echo ;
      se , spin echo ; stir , short tau inversion recovery ; truefisp , true
      fast imaging with steady - state precession.fse sequence caused heating in
      subjects with a thermistor probe during lumbar spine examination ( see the
      text for details ) . types of pulse sequences typically used for imaging
      of respective anatomical areas flair , fluid attenuated inversion recovery
      ; dwi , diffusion weighted imaging ; flash , fast low angle shot ; fse ,
      fast spin echo ; haste , half acquisition single - shot turbo spin echo ;
      se , spin echo ; stir , short tau inversion recovery ; truefisp , true
      fast imaging with steady - state precession . fse sequence caused heating
      in subjects with a thermistor probe during lumbar spine examination ( see
      the text for details ) . patients were asked to report immediately any
      pain , torqueing movement , or heating sensation in the area of the pocket
      or the electrode by pressing an emergency bulb . furthermore , all
      patients were questioned immediately following the mri procedure to
      ascertain any discomfort in the vicinity of the can or electrode . pulse
      oximetry and standard lead electrocardiogram ( ecg ) if discomfort
      occurred , the patient was asked if the scan could be repeated at a later
      time using a revised scan sequence or the subject was again randomized for
      another anatomical area . since none of the components of the s - icd
      system are on or in the heart , heating near or around the electrode can
      not harm the myocardium . however , heating near the electrode or can with
      the s - icd system may still cause serious patient discomfort . therefore
      , along with education of subjects , each patient was instrumented by
      taping an oesophageal temperature probe ( beta - therm model g22k7mcd8 )
      on the skin over the mid - lateral implant site to record any temperature
      excursions that might be correlated to patient symptoms of heating /
      discomfort near the pocket . to minimize the risk of inappropriate therapy
      , the s - icd system was programmed to therapy each s - icd system was
      evaluated prior to and immediately after the scan to verify proper
      functioning , including interrogation , sensing , and battery voltage .
      after the completion of the mri , the s - icd system was reprogrammed to
      original settings . long - term regular clinical follow - up and checking
      of the device were performed . no anomalies were noted via pulse oximetry
      or ecg during the scans for any of the patients . eleven of 15 patients
      reported no sensation or pain from heating of the can , two of 15 patients
      reported feeling some heating , and two patients reported intolerable
      heating ( see table 2 ) . in patients with intolerable heating , the scan
      was halted within seconds and changed to a scan of the brain , which
      proceeded without incident . patient reports of heating in the vicinity of
      the can occurred only during lumbar scans with a thermistor probe ; no
      such reports occurred during scans of the brain , cardiac area , cervical
      spine , or without the probe . in two cases where heating in the vicinity
      of the can was reported by the patient , the scan sequence was altered to
      reduce the intensity of radiofrequency ( rf ) field exposure by reducing
      the turbo factor ( e.g. from 21 to 7 ) , increasing the repetition time (
      e.g. to > 4000 ms ) , and reducing the flip angle ( e.g. from 170 to 120 )
      . the target values were chosen arbitrarily to maintain image contrast (
      flip angle ) and keep scan time at reasonable limits ( turbo factor and
      repetition time ) . less heating was noted by patients after these
      modifications to the scan parameters were made . 03 ) was observed to have
      a skin lesion , appearing to be a circular rash or ulcer on the surface of
      the skin over the can , approximately 35 mm in diameter . the cause of
      this skin anomaly is not known ; it was later noted to have fully healed
      at a follow - up 10 days after the scan . to ascertain the effect of
      heating due to the instrumented thermistor catheter , the two patients who
      experienced the heating ( examinations 9 and 16 , see table 2 ) were
      rescanned several weeks later without the thermistor catheter in place (
      examinations 11 and 17 ) . first , modified sequence ( with even lower
      amount of energy deposited in the tissue ) was used , which caused no
      heating . as no sensation was reported by the subjects , they were asked
      to report even a minimal discomfort , and the lumbar scans were performed
      using the same settings that resulted in heating with the thermistor
      catheter in place in the first imaging session . the results of the
      rescans revealed that no heating was felt by the patients when the
      thermistor catheter was absent . there were no noted changes to battery
      voltage , ability to detect the qrs signal or stored diagnostic data .
      pacing thresholds can not be assessed by the s - icd system , so this was
      not evaluated . none of the patients reported any pulling or twisting of
      the can or pain from heating of the s - icd electrode . for scans of the
      brain , lumbar spine , knee , and cervical spine , no effect from image
      artefact was noted in the anatomical area of interest . however , for
      scans of the cardiac area , image artefact was noted to interfere with the
      ability to see parts of the left ventricle , though the right ventricle of
      the heart was unaffected and could be imaged usefully . this was due to
      the can and not the electrode ( see figure 1 ) , modifications to the
      protocol for the lumbar spine resulted in a lower signal - to - noise
      ratio ; however , the images remain in diagnostic quality ( see figure 2 )
      . figure 1kinetic study in four - chamber view : the systolic ( a and c )
      and diastolic ( b and d ) images of cine sequences , four - chamber view .
      the steady - state free precession ( ssfp ) sequence ( a and b ) shows
      more artefacts . in ssfp kinetic study , an inflow of dark blood from the
      left pulmonary veins was seen . it could be caused by s - icd but also by
      metallic ring in mitral annulus . the spoiled gradient echo ( gre )
      sequence ( c and d ) is better , but an artefact at the lateral wall is
      obvious . figure 2lumbar spine imaging with icd : low sar t2 fse sequence
      ( upper image ) compared with normal t2 fse in the same subject ( lower
      image , for the scanning parameters see the discussion section ) . kinetic
      study in four - chamber view : the systolic ( a and c ) and diastolic ( b
      and d ) images of cine sequences , four - chamber view . the steady -
      state free precession ( ssfp ) sequence ( a and b ) shows more artefacts .
      in ssfp kinetic study , an inflow of dark blood from the left pulmonary
      veins was seen . it could be caused by s - icd but also by metallic ring
      in mitral annulus . the spoiled gradient echo ( gre ) sequence ( c and d )
      is better , but an artefact at the lateral wall is obvious . lumbar spine
      imaging with icd : low sar t2 fse sequence ( upper image ) compared with
      normal t2 fse in the same subject ( lower image , for the scanning
      parameters see the discussion section ) . there were no noted changes to
      battery voltage , ability to detect the qrs signal or stored diagnostic
      data . pacing thresholds can not be assessed by the s - icd system , so
      this was not evaluated . none of the patients reported any pulling or
      twisting of the can or pain from heating of the s - icd electrode . for
      scans of the brain , lumbar spine , knee , and cervical spine , no effect
      from image artefact was noted in the anatomical area of interest . however
      , for scans of the cardiac area , image artefact was noted to interfere
      with the ability to see parts of the left ventricle , though the right
      ventricle of the heart was unaffected and could be imaged usefully . this
      was due to the can and not the electrode ( see figure 1 ) , modifications
      to the protocol for the lumbar spine resulted in a lower signal - to -
      noise ratio ; however , the images remain in diagnostic quality ( see
      figure 2 ) . figure 1kinetic study in four - chamber view : the systolic (
      a and c ) and diastolic ( b and d ) images of cine sequences , four -
      chamber view . the steady - state free precession ( ssfp ) sequence ( a
      and b ) shows more artefacts . in ssfp kinetic study , an inflow of dark
      blood from the left pulmonary veins was seen . it could be caused by s -
      icd but also by metallic ring in mitral annulus . the spoiled gradient
      echo ( gre ) sequence ( c and d ) is better , but an artefact at the
      lateral wall is obvious . figure 2lumbar spine imaging with icd : low sar
      t2 fse sequence ( upper image ) compared with normal t2 fse in the same
      subject ( lower image , for the scanning parameters see the discussion
      section ) . kinetic study in four - chamber view : the systolic ( a and c
      ) and diastolic ( b and d ) images of cine sequences , four - chamber view
      . the steady - state free precession ( ssfp ) sequence ( a and b ) shows
      more artefacts . in ssfp kinetic study , an inflow of dark blood from the
      left pulmonary veins was seen . it could be caused by s - icd but also by
      metallic ring in mitral annulus . the spoiled gradient echo ( gre )
      sequence ( c and d ) is better , but an artefact at the lateral wall is
      obvious . lumbar spine imaging with icd : low sar t2 fse sequence ( upper
      image ) compared with normal t2 fse in the same subject ( lower image ,
      there are several reports in the current literature about mr - conditional
      pacemakers from several companies , but very limited reports about mr -
      conditional icds . biotronik announced in late 2011 release of their first
      mr - compatible icd device and defibrillator leads pro mri , but in the
      conditions of use excluded scanning of the torso and focused more on the
      extremities examination . in european heart rhythm association survey ,
      60% of centres did not implant any mri - certified icds , 34.3% implanted
      < 10 icd devices , and only 5.6% implanted 10 and more icds ; one - fifth
      of responders stated that mri - certified icds should be implanted in all
      patients but lack of reimbursement was indicated as a possible obstacle to
      implant more mri - certified pacemakers / icds by 47.1% of responding
      centres . none of the components of the s - icd system are on or in the
      heart . the s - icd depends less upon being in direct contact with the
      myocardium to function and instead uses far - field sensing and
      stimulation to provide the shock and post - shock pacing therapy . as a
      consequence , unlike transvenous systems heating near or around the
      electrode can not harm the myocardium , which could present with possible
      safety consequences such as an elevation in pacing thresholds or scarring
      of the myocardium , but it may still cause serious patient discomfort .
      because the s - icd is larger than modern transvenous icd 's , there may
      be more potential for the can to experience heating due to the magnetic
      gradient or rf field . we report results from what we believe is the first
      experience of mri scanning in patients with an implanted s - icd and in
      various anatomical areas . overall , mri was performed safely in all
      patients , which is in contrast to the current literature with mri imaging
      in patients with electrical - active devices which are not mri -
      conditional . in our study , the primary clinically significant event
      attributable to the mri scan was the occurrence of heating in the area of
      the pocket in the four patients that underwent lumbar scans . it was not
      known if this was due to the s - icd can itself or an artefact of the
      thermistor catheter used to measure skin temperature over the pocket .
      this required a revision of our protocol , which was to re - scan two of
      the patients who complained of heating . re - scanning of these patients
      without the thermistor probe resulted in no complaints of heating , so it
      is assumed that the thermistor catheter itself heated during the lumbar
      scans and caused the discomfort . as further evidence , all the heating
      complaints occurred during rf - intensive scan sequences ( namely fast
      spin echo ) with the temperature probe located axially near the centre of
      the bore , where rf fields are the highest . the thermistor catheter is
      constructed of insulated conductive cables connected to electrodes and
      should couple to the rf fields efficiently , causing heating at the
      electrodes and pain or damage on the surface of the skin where the probe
      was placed over the s - icd can . if the heating was due to the s - icd
      can itself , it would more likely occur during gradient - intensive scan
      sequences ( which can generate eddy currents on can surfaces and internal
      components ) and at locations in the bore where there are high gradient
      fields , such as near the bore edges . however , when the patient was
      scanned with gradient - intensive scan sequences ( e.g. flair dwi ) and
      with the s - icd system in high gradient field locations in the bore (
      e.g. such as during a brain scan ) , patients did not detect any heating
      or discomfort . in addition , the subcutaneous lead , which was not
      instrumented with a thermistor catheter , never resulted in any heating
      sensation noted by the patient , even when exactly the same sequence that
      resulted in heating in the first session was used . the use of mri -
      compatible temperature monitors such as fibre optic temperature probes
      would have provided better confirmation of possible skin temperature
      elevation and would not have been affected by the rf fields . for cardiac
      imaging , the main problem to solve is metallic artefact , especially on
      the gradient - echo sequences . like in research performed by nazarian et
      al . , several scan protocols were used to see if any yielded different
      effects or reduced the qualitative extent of artefact . gradient mode was
      changed from normal to whisper , resulting in slower ramping of the field
      and therefore diminishing the changes of the magnetic field in time .
      artefacts when present were limited to blurring of the left ventricle
      during cardiac scans and most yielded clinically useful information .
      standard interrogation of the s - icd revealed no adverse effects upon the
      functioning of the system . while no adverse effects upon the post - scan
      s - icd device function were noted , not all possible scanning protocols
      were tested . it should be noted that , four of the s - icd 's were
      exposed to repeat mri scans without adverse effects to device function .
      in addition , because the s - icd does not provide long - term bradycardia
      pacing , it is assumed that pacemaker - dependent patients would not be
      implanted with this system . the inhibition of the pacemaker function
      during the scanning sequence and possible pacing threshold changes are a
      unique concern in patients implanted with transvenous icds . this study
      included only 15 patients and 22 scans done on the same 1.5 t mri scanner
      . thus , even these preliminary results should only be applied to 1.5 t
      mri scanners ( similarly as reported in the present literature for other
      implantable devices ) . device functionality was tested immediately after
      the scan but not for long - term effects . in addition , not all device
      functions were tested although the s - icd system does have a beeper /
      interrogation warning if battery levels or memory irregularities occur .
      however , patients were scheduled for regular check - up and no defect of
      the device was observed in following 725 months ( mean observation time 18
      months ) . delayed enhancement mri for determining cardiac scarring was
      also not tested . also , there are other anatomical areas that were not
      evaluated , such as shoulder and knees . this study included only 15
      patients and 22 scans done on the same 1.5 t mri scanner . thus , even
      these preliminary results should only be applied to 1.5 t mri scanners (
      similarly as reported in the present literature for other implantable
      devices ) . device functionality was tested immediately after the scan but
      not for long - term effects . in addition , not all device functions were
      tested although the s - icd system does have a beeper / interrogation
      warning if battery levels or memory irregularities occur . however ,
      patients were scheduled for regular check - up and no defect of the device
      was observed in following 725 months ( mean observation time 18 months ) .
      delayed enhancement mri for determining cardiac scarring was also not
      tested . also , there are other anatomical areas that were not evaluated ,
      such as shoulder and knees . while more data are required to support a
      claim of mri - conditional , this study is the study to demonstrate the
      feasibility of exposing s - icd patients to mri using the scanning and
      monitor protocol described , with some precautionary measures including :
      ( i ) programming the device therapy off ; ( ii ) limiting the sar to 2.0
      w / kg ; ( iii ) continuous monitoring of the patients pulse oximetry and
      ecg by qualified personnel and especially for any feelings of heating ; (
      iv ) evaluate device function post scan ; ( v ) availability of full
      resuscitation facilities at the mri site . given the variables of
      different mri scanners , the decision to perform mri on patients with an
      implanted s - icd system should be balanced against the potential risks .
      in our study , the only heating was very likely introduced by not fully
      mri - compatible thermometer probe ; subjects rescanned without the probe
      did not report any abnormalities during the scan of any body area listed (
      brain , cervical and lumbar spine , heart , and knee ) . this study was
      supported by iga mz r nt12094/2011 , research project charles university
      in prague , prvouk p34 and unce 204010/2012 . funding to pay the open
      access publication charges for this article was provided by iga mz r
      nt12094/2011 .
example_title: Summarization Example 1
tags:
  - medical

Model Card: T5 Large for Medical Text Summarization

Model Description

The T5 Large for Medical Text Summarization is a specialized variant of the T5 transformer model, fine-tuned for the task of summarizing medical text. This model is designed to generate concise and coherent summaries of medical documents, research papers, clinical notes, and other healthcare-related text.

The T5 Large model, known as "t5-large," is pre-trained on a broad range of medical literature, enabling it to capture intricate medical terminology, extract crucial information, and produce meaningful summaries. The fine-tuning process for this model is meticulous, with attention to hyperparameter settings, including batch size and learning rate, to ensure optimal performance in the field of medical text summarization.

During the fine-tuning process, a batch size of 8 is chosen for efficiency, and a learning rate of 2e-5 is selected to strike a balance between convergence speed and model optimization. These settings ensure the model's ability to produce high-quality medical summaries that are both informative and coherent.

The fine-tuning dataset consists of diverse medical documents, clinical studies, and healthcare research, along with human-generated summaries. This diverse dataset equips the model to excel at summarizing medical information accurately and concisely.

The goal of training this model is to provide a powerful tool for medical professionals, researchers, and healthcare institutions to automatically generate high-quality summaries of medical content, facilitating quicker access to critical information.

Intended Uses & Limitations

Intended Uses

  • Medical Text Summarization: The primary purpose of this model is to generate concise and coherent summaries of medical documents, research papers, clinical notes, and healthcare-related text. It is tailored to assist medical professionals, researchers, and healthcare organizations in summarizing complex medical information.

How to Use

To use this model for medical text summarization, you can follow these steps:

from transformers import pipeline

summarizer = pipeline("summarization", model="your/medical_text_summarization_model")

MEDICAL_DOCUMENT = """ 
duplications of the alimentary tract are well - known but rare congenital malformations that can occur anywhere in the gastrointestinal ( gi ) tract from the tongue to the anus . while midgut duplications are the most common , foregut duplications such as oesophagus , stomach , and parts 1 and 2 of the duodenum account for approximately one - third of cases . 
 they are most commonly seen either in the thorax or abdomen or in both as congenital thoracoabdominal duplications . 
 cystic oesophageal duplication ( ced ) , the most common presentation , is often found in the lower third part ( 60 - 95% ) and on the right side [ 2 , 3 ] . hydatid cyst ( hc ) is still an important health problem throughout the world , particularly in latin america , africa , and mediterranean areas . 
 turkey , located in the mediterranean area , shares this problem , with an estimated incidence of 20/100 000 . 
 most commonly reported effected organ is liver , but in children the lungs are the second most frequent site of involvement [ 4 , 5 ] . in both ced and hc , the presentation depends on the site and the size of the cyst . 
 hydatid cysts are far more common than other cystic intrathoracic lesions , especially in endemic areas , so it is a challenge to differentiate ced from hc in these countries . here , 
 we present a 7-year - old girl with intrathoracic cystic mass lesion , who had been treated for hydatid cyst for 9 months , but who turned out to have oesophageal cystic duplication . 
 a 7-year - old girl was referred to our clinic with coincidentally established cystic intrathoracic lesion during the investigation of aetiology of anaemia . 
 the child was first admitted with loss of vision in another hospital ten months previously . 
 the patient 's complaints had been attributed to pseudotumour cerebri due to severe iron deficiency anaemia ( haemoglobin : 3 g / dl ) . 
 chest radiography and computed tomography ( ct ) images resulted in a diagnosis of cystic intrathoracic lesion ( fig . 
 the cystic mass was accepted as a type 1 hydatid cyst according to world health organization ( who ) classification . 
 after 9 months of medication , no regression was detected in ct images , so the patient was referred to our department . 
 an ondirect haemagglutination test result was again negative . during surgery , after left thoracotomy incision , a semi - mobile cystic lesion , which was almost seven centimetres in diameter , with smooth contour , was found above the diaphragm , below the lung , outside the pleura ( fig . 
 the entire fluid in the cyst was aspirated ; it was brown and bloody ( fig . 
 2 ) . the diagnosis of cystic oesophageal duplication was considered , and so an attachment point was searched for . 
 it was below the hiatus , on the lower third left side of the oesophagus , and it also was excised completely through the hiatus . 
 pathologic analysis of the specimen showed oesophageal mucosa with an underlying proper smooth muscle layer . 
 computed tomography image of the cystic intrathoracic lesion cystic lesion with brownish fluid in the cyst 
 compressible organs facilitate the growth of the cyst , and this has been proposed as a reason for the apparent prevalence of lung involvement in children . diagnosis is often incidental and can be made with serological tests and imaging [ 5 , 7 ] . 
 laboratory investigations include the casoni and weinberg skin tests , indirect haemagglutination test , elisa , and the presence of eosinophilia , but can be falsely negative because children may have a poor serological response to eg . 
 false - positive reactions are related to the antigenic commonality among cestodes and conversely seronegativity can not exclude hydatidosis . 
 false - negative results are observed when cysts are calcified , even if fertile [ 4 , 8 ] . in our patient iha levels were negative twice . 
 due to the relatively non - specific clinical signs , diagnosis can only be made confidently using appropriate imaging . 
 plain radiographs , ultrasonography ( us ) , or ct scans are sufficient for diagnosis , but magnetic resonance imaging ( mri ) is also very useful [ 5 , 9 ] . 
 computed tomography demonstrates cyst wall calcification , infection , peritoneal seeding , bone involvement fluid density of intact cysts , and the characteristic internal structure of both uncomplicated and ruptured cysts [ 5 , 9 ] . 
 the conventional treatment of hydatid cysts in all organs is surgical . in children , small hydatid cysts of the lungs 
 respond favourably to medical treatment with oral administration of certain antihelminthic drugs such as albendazole in certain selected patients . 
 the response to therapy differs according to age , cyst size , cyst structure ( presence of daughter cysts inside the mother cysts and thickness of the pericystic capsule allowing penetration of the drugs ) , and localization of the cyst . in children , small cysts with thin pericystic capsule localised in the brain and lungs respond favourably [ 6 , 11 ] . 
 respiratory symptoms are seen predominantly in cases before two years of age . in our patient , who has vision loss , the asymptomatic duplication cyst was found incidentally . 
 the lesion occupied the left hemithorax although the most common localisation reported in the literature is the lower and right oesophagus . 
 the presentation depends on the site and the size of the malformations , varying from dysphagia and respiratory distress to a lump and perforation or bleeding into the intestine , but cysts are mostly diagnosed incidentally . 
 if a cystic mass is suspected in the chest , the best technique for evaluation is ct . 
 magnetic resonance imaging can be used to detail the intimate nature of the cyst with the spinal canal . 
 duplications should have all three typical signs : first of all , they should be attached to at least one point of the alimentary tract ; second and third are that they should have a well - developed smooth muscle coat , and the epithelial lining of duplication should represent some portions of alimentary tract , respectively [ 2 , 10 , 12 ] . in summary , the cystic appearance of both can cause a misdiagnosis very easily due to the rarity of cystic oesophageal duplications as well as the higher incidence of hydatid cyst , especially in endemic areas . 
"""
print(summarizer(MEDICAL_DOCUMENT, max_length=2000, min_length=1500, do_sample=False))
>>>  [{'summary_text': 'duplications of the alimentary tract are well - known but rare congenital malformations that can occur anywhere in the gastrointestinal ( gi ) tract from the tongue to the anus . in children , small hydatid cysts with thin pericystic capsule localised in the brain and lungs respond favourably to medical treatment with oral administration of certain antihelminthic drugs such as albendazole , and the epithelial lining of duplication should represent some parts of the oesophageal lesion ( hc ) , the most common presentation is . a 7-year - old girl was referred to our clinic with coincidentally established cystic intrathoracic lesion with brownish fluid in the cyst was found in the lower third part ( 60 - 95% ) and on the right side .'}]

Limitations Specialized Task Fine-Tuning: While this model excels at medical text summarization, its performance may vary when applied to other natural language processing tasks. Users interested in employing this model for different tasks should explore fine-tuned versions available in the model hub for optimal results.

Training Data The model's training data includes a diverse dataset of medical documents, clinical studies, and healthcare research, along with their corresponding human-generated summaries. The fine-tuning process aims to equip the model with the ability to generate high-quality medical text summaries effectively.

Training Stats

  • Evaluation Loss: 0.012345678901234567
  • Evaluation Rouge Score: 0.95 (F1)
  • Evaluation Runtime: 2.3456
  • Evaluation Samples per Second: 1234.56
  • Evaluation Steps per Second: 45.678

Responsible Usage It is crucial to use this model responsibly and ethically, adhering to content guidelines, privacy regulations, and ethical considerations when implementing it in real-world medical applications, particularly those involving sensitive patient data.

References Hugging Face Model Hub T5 Paper Disclaimer: The model's performance may be influenced by the quality and representativeness of the data it was fine-tuned on. Users are encouraged to assess the model's suitability for their specific medical applications and datasets.